Team:DTU-Denmark/Methods/Calibrating Electrodes
From 2013.igem.org
(Difference between revisions)
(→Calibration of NO and N2O electrodes/probes) |
(→Calibration of NO and N2O electrodes/probes) |
||
Line 13: | Line 13: | ||
# Make standard solution NO | # Make standard solution NO | ||
- | + | ## Make 200 ml solution 1 M NaOH | |
- | + | ## Make it anoxic by bubbling with N<sub>2</sub> for 5 minutes | |
- | + | ## Make MQ water anoxic | |
- | + | ## Bubble 5% NO solution through NaOH solution | |
- | + | ## Convey towards the water flask for 20 minutes | |
# Make standard solution of N<sub>2</sub>O | # Make standard solution of N<sub>2</sub>O | ||
- | + | ## Make solution anoxic by bubbling 5 minutes with N<sub>2</sub> | |
- | + | ## Flush with N<sub>2</sub>O for 20 minutes. | |
*Procedure | *Procedure | ||
Line 31: | Line 31: | ||
# Start by calibrating N<sub>2</sub>O sensor (the adding points are the concentration of N<sub>2</sub>O in mg/L) | # Start by calibrating N<sub>2</sub>O sensor (the adding points are the concentration of N<sub>2</sub>O in mg/L) | ||
- | + | ## Add the first point “0” on the sensor | |
- | + | ## Spike 0.5 ml of N<sub>2</sub>O in 200 ml and add point “60.35” | |
- | + | ## Spike 0.5 ml of N<sub>2</sub>O and add point “120.7” | |
- | + | ## Spike 0.5 ml of N<sub>2</sub>O and add point “181.06” | |
- | + | ## Spike 0.5 ml of N<sub>2</sub>O and add point “2241.41” | |
- | + | ## Spike 1 ml of N<sub>2</sub>O and add point “362.11” | |
- | + | ## Spike 1 ml of N<sub>2</sub>O and add point “482.81” | |
- | + | ## Then apply the calibration for the sensor | |
- | + | # Continue by calibrating NO sensor (the adding points are the concentration of NO in mg/L) | |
- | + | ## Add the first point “0” on the sensor | |
- | + | ## Spike 1 ml of NO in 200 ml and add point “0.47” | |
- | + | ## Spike 1 ml of NO and add point “0.94” | |
- | + | ## Spike 1 ml of NO and add point “1.4” | |
- | + | ## Spike 1 ml of NO and add point “1.86” | |
- | + | ## Spike 1 ml of NO and add point “2.31” | |
- | + | ## Then apply the calibration for the sensor | |
- | + | ||
- | + | ||
- | + | ||
- | + | ||
- | + | ||
- | + | ||
- | + | ||
- | + | ||
{{:Team:DTU-Denmark/Templates/Footer1}} | {{:Team:DTU-Denmark/Templates/Footer1}} |
Revision as of 17:33, 12 August 2013
Calibration of electrodes
Calibration of NO and N2O electrodes/probes
- Cleaning the probes
- Prepare 96% ethanol solution
- Prepare 0.01M HCl solution
- Put both probes in ethanol solution for 1 minute
- Put both probes in HCl solution for 1 minute
- Rinse with demineralized water
- Calibration
- Make standard solution NO
- Make 200 ml solution 1 M NaOH
- Make it anoxic by bubbling with N2 for 5 minutes
- Make MQ water anoxic
- Bubble 5% NO solution through NaOH solution
- Convey towards the water flask for 20 minutes
- Make standard solution of N2O
- Make solution anoxic by bubbling 5 minutes with N2
- Flush with N2O for 20 minutes.
- Procedure
- Pure 200 ml of DM minimal medium in the bottle
- Adjust the temperature at around 37oC by using a magnetic stirrer
- Adjust both of the probes in the lid of the bottle
- Connect the electrodes with the microsensors
- Make the solution anoxic by bubbling with N2 for 10 minutes
- Start by calibrating N2O sensor (the adding points are the concentration of N2O in mg/L)
- Add the first point “0” on the sensor
- Spike 0.5 ml of N2O in 200 ml and add point “60.35”
- Spike 0.5 ml of N2O and add point “120.7”
- Spike 0.5 ml of N2O and add point “181.06”
- Spike 0.5 ml of N2O and add point “2241.41”
- Spike 1 ml of N2O and add point “362.11”
- Spike 1 ml of N2O and add point “482.81”
- Then apply the calibration for the sensor
- Continue by calibrating NO sensor (the adding points are the concentration of NO in mg/L)
- Add the first point “0” on the sensor
- Spike 1 ml of NO in 200 ml and add point “0.47”
- Spike 1 ml of NO and add point “0.94”
- Spike 1 ml of NO and add point “1.4”
- Spike 1 ml of NO and add point “1.86”
- Spike 1 ml of NO and add point “2.31”
- Then apply the calibration for the sensor