Plasmid purification

Protocole from Macherey-Nagel plasmid purification notebook

. Récupération des bactéries
Set saturated E.coli LM culture into 2 mL tubes. Centrifuge at 11 000 x g for 30 secondes. Discard as much as supernatant as possible.



.Cell lysis
Add 250 μL Buffer A1 (resuspension buffer). Resuspend the cells with a vortex or a pipette.

Add 250 μL Buffer A2 (lysis buffer). Mix gently by inverting the tube 6 - 8 times. Incubate at room temperature until lysate appears clear.

Add 300 μL Buffer A3 (neutralisation buffer). Mix thoroughly by inverting the tube 6 - 8 times Mélanger doucement en retournant les tubes 6 à 8 fois.

. Clarification of lysate
Centrifuge at 11 000 x g for 5 minutes . Repeat this step until supernatant is not clear.

.Bind ADN
Place a NucleoSpin Plasmid Column in a Collection Tube of 2 mL et and set the supernatant from the last step. Centrifuge at 11 000 x g for 1 minute. Discard flow-through and place the column back into the collection tube.

. Wash membrane
Add 600 μL Buffer A4 (wash buffer) previously supplemented with ethanol. Centrifuge ar 11 000 x g for 1 minute. Discard flow-through and place the column back into an empty collection tube.
. Dry membrane
Centrifuge at 11 000 x g for 2 minutes and discard the collection tube.

. Elution de l'ADN
Place the column in a 1,5 mL and add 50 μL de Buffer AE (elution buffer). Incubate at room temperature and centrifuge at 11 000 x g for 1 minute.
Measure the concentration with nanodrop, then stock the tubes at -20°C