<h4>Temperature</h4>
+
<div class="twelve columns">
-
<img src="https://static.igem.org/mediawiki/2013/3/33/800px-Temptest.jpg" alt="heat&cool" style="max-height: 300px margin-left: 100px">
+
 
 +
<h4>Temperature</h4>
 +
</div>
 +
<div class="six columns" style="margin-bottom:0px">
 +
With the temperature sensor and heating circuit placed inside the incubation chamber, the rate of heating when full power (100% PWM) was applied and the rate of cooling when the heating circuit was switched off was graphed in Figure 1. Some initial calibration runs were then performed as we fine tuned our heating algorithm.
 +
</div>
 +
<div class="six columns">
 +
<img src="https://static.igem.org/mediawiki/2013/3/33/800px-Temptest.jpg" alt="heat&cool" style="margin-top:0px; max-height:none;">
 +
</div>
 +
<div class="twelve columns">
 +
<img src="https://static.igem.org/mediawiki/2013/2/23/Fig1heat%26cool.png" alt="heat&cool" style="max-height:none; margin-top:0px;">
 +
<div style="margin-top:-20px" class="center"> Figure 1: Heating and Cooling curves</div>  
<br>
<br>
-
With the temperature sensor and heating circuit placed inside the incubation chamber, the rate of heating when full power (100% PWM) was applied and the rate of cooling when the heating circuit was switched off was graphed in Figure 1. Some initial calibration runs were then performed as we fine tuned our heating algorithm.
+
As the optimal temperature for fungi growth is 27°C, we calculated the difference between the measured temperature and optimal temperature. When the difference is more than 5°C cooler than optimal ie. more than 25°C, maximum heat is applied so as to quickly heat the chamber. As the temperature difference decreases, the duty cycle decreases proportionally until the temperature is at 27°C. Once the temperature increases beyond the optimal, the heating circuit shuts off.
-
 
+
-
<img src="https://static.igem.org/mediawiki/2013/2/23/Fig1heat%26cool.png" alt="heat&cool" style="max-height:none">
+
-
<div style="margin-top:-20px" class="center"> Figure 1: Heating and Cooling curves</div> <br>
+
-
 
+
-
As the optimal temperature for fungi growth is 27°C, we calculated the difference between the measured temperature and optimal temperature. When the difference is more than 10°C cooler than optimal ie. 17°C, maximum heat is applied so as to quickly heat the chamber. As the temperature difference decreases, the duty cycle decreases proportionally until the temperature is at 27°C. Once the temperature increases beyond the optimal, the heating circuit shuts off.
+
-
 
+
<img src="https://static.igem.org/mediawiki/2013/b/b7/Figure2optimaltemperaturecalibration.png" alt="temperaturecalibration" style="max-height:none">  
<img src="https://static.igem.org/mediawiki/2013/b/b7/Figure2optimaltemperaturecalibration.png" alt="temperaturecalibration" style="max-height:none">  
-
<div class="center" style="margin-top:-20px">Figure 2: Calibration to Optimal Temperature</div> <br>
+
<div class="center" style="margin-top:-20px">Figure 2: Calibration to Optimal Temperature</div>  
-
 
+
<br>
As seen in Figure 2, From room temperature (24°C), it takes less than 5 minutes for the incubation chamber to reach optimal temperature (27°C) and it remains stable.
As seen in Figure 2, From room temperature (24°C), it takes less than 5 minutes for the incubation chamber to reach optimal temperature (27°C) and it remains stable.
-
 
+
<br>
-
<h4>Humidity</font></h4>
+
<h4>Humidity</h4>
A HH10D humidity sensor module was used to measure the relative humidity of the environment.  This sensor was connected according to manufacturer instructions such as connecting power, ground, frequency output, SCL, and SDA to the corresponding pins on the sensor board.<br><br>   
A HH10D humidity sensor module was used to measure the relative humidity of the environment.  This sensor was connected according to manufacturer instructions such as connecting power, ground, frequency output, SCL, and SDA to the corresponding pins on the sensor board.<br><br>   
To collect data about the humidity, a mistmaker was used to increase the relative humidity of the surrounding area, and the humidity sensor was placed in the vicinity of this mistmaker from the start.  This experiment was done by putting the mistmaker in a box, making sure to not allow the mist to escape, and as the mistmaker distributed the water more evenly throughout the box, the relative humidity increased.  As the amount of humidity began to increase, the sensor could measure the increase in relative humidity.
To collect data about the humidity, a mistmaker was used to increase the relative humidity of the surrounding area, and the humidity sensor was placed in the vicinity of this mistmaker from the start.  This experiment was done by putting the mistmaker in a box, making sure to not allow the mist to escape, and as the mistmaker distributed the water more evenly throughout the box, the relative humidity increased.  As the amount of humidity began to increase, the sensor could measure the increase in relative humidity.
-
 
+
<br>
-
<img src="https://static.igem.org/mediawiki/2013/4/46/Humiditygraph.jpg" alt="humidity" style="max-height:none">
+
<center><img src="https://static.igem.org/mediawiki/2013/4/46/Humiditygraph.jpg" alt="humidity" style="max-height:300px" align="middle">
-
<div class="center" style="margin-top:-20px">Figure 3: Increase in humidity levels using a mistmaker</div>  
+
<div class="center" style="margin-top:-20px">Figure 3: Increase in humidity levels using a mistmaker</div> </center>
 +
<div>
</div>
</div>
</div>
</div>

Latest revision as of 03:06, 29 October 2013

Cornell University Genetically Engineered Machines

Results

Temperature

With the temperature sensor and heating circuit placed inside the incubation chamber, the rate of heating when full power (100% PWM) was applied and the rate of cooling when the heating circuit was switched off was graphed in Figure 1. Some initial calibration runs were then performed as we fine tuned our heating algorithm.
heat&cool
heat&cool
Figure 1: Heating and Cooling curves

As the optimal temperature for fungi growth is 27°C, we calculated the difference between the measured temperature and optimal temperature. When the difference is more than 5°C cooler than optimal ie. more than 25°C, maximum heat is applied so as to quickly heat the chamber. As the temperature difference decreases, the duty cycle decreases proportionally until the temperature is at 27°C. Once the temperature increases beyond the optimal, the heating circuit shuts off. temperaturecalibration
Figure 2: Calibration to Optimal Temperature

As seen in Figure 2, From room temperature (24°C), it takes less than 5 minutes for the incubation chamber to reach optimal temperature (27°C) and it remains stable.

Humidity

A HH10D humidity sensor module was used to measure the relative humidity of the environment. This sensor was connected according to manufacturer instructions such as connecting power, ground, frequency output, SCL, and SDA to the corresponding pins on the sensor board.

To collect data about the humidity, a mistmaker was used to increase the relative humidity of the surrounding area, and the humidity sensor was placed in the vicinity of this mistmaker from the start. This experiment was done by putting the mistmaker in a box, making sure to not allow the mist to escape, and as the mistmaker distributed the water more evenly throughout the box, the relative humidity increased. As the amount of humidity began to increase, the sensor could measure the increase in relative humidity.
humidity
Figure 3: Increase in humidity levels using a mistmaker